FDA Adverse Event Malfunction Summary report: N

GREENLIGHT MOXY FIBER OPTIC

MDR report key: 3946183 · Received July 21, 2014

Report

Report Number
2937094-2014-00639
Event Type
Malfunction
Date Received
July 21, 2014
Date of Event
July 9, 2014
Report Date
July 10, 2014
Manufacturer
AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
Product Code
GEX
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVALUATION: FAILURE ANALYSIS FOR FIBER (B)(4): THE FIBER SHOWS A CIRCUMFERENTIAL FRACTURE DISTAL TO FIBER/CAP FUSION ZONE DISTAL TO THE BEVEL EDGE; THE FIBER PROXIMAL TO FRACTURE CAN ROTATE INDEPENDENTLY OF METAL CAP; THE GLASS CAP EXHIBITS SEVERE DEVITRIFICATION; THE METAL CAP EXHIBITS MODERATE DETRITUS ADHESION. BASED ON THE ANALYSIS ABOVE, THE POTENTIAL FOR FORWARD FIRING MAY EXIST. PROBABLE ROOT CAUSE: BASED ON THE PRODUCT ANALYSIS RESULTS, THE PROBABLE ROOT CAUSE OF THE FAILURE IS: HEAT ACCUMULATION. CAP WEAR WAS ACCELERATED DUE TO ANATOMICAL/PROCEDURAL FACTORS (TISSUE CONTACT AND TECHNIQUE) ENCOUNTERED DURING THE PROCEDURE WHICH WOULD LIMIT THE PERFORMANCE OF THE FIBER.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PROSTATE PROCEDURE, THE FIRST FIBER ISSUE ¿THE FIBER STOPPED ROTATING, LIKE THE INNER AND OUTER SHEATH BECAME SEPARATED". AT 191,039 JOULES. THE FIBER WAS EXCHANGED; THE SECOND FIBER ISSUE ¿AIM BEAM FIRES STRAIGHT OUT OF THE FIBER" AT 39,657 JOULES. THE FIBER WAS EXCHANGED; THE THIRD FIBER ISSUE ¿AIM BEAM FIRES STRAIGHT OUT OF THE FIBER" AT 50,559 JOULES. THE FIBER WAS EXCHANGED, THE FOURTH FIBER ISSUE "AIM BEAM FIRES STRAIGHT OUT OF THE FIBER" AT 74,296 JOULES. THIS FIBER WAS ALSO USED TO COMPLETE THE CASE. PATIENT OUTCOME: NO PATIENT OR USER INJURY REPORTED. THIS REPORT IS FOR THE 3RD FIBER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
424758 GREENLIGHT MOXY FIBER OPTIC POWERED LASER SURGICAL INSTRUMENT GEX AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE) 0010-2400 401A

Patients

Seq Age Sex Outcome Treatment
1