34KHZ SM LONG ANGLED W/O TIP
Report
- Report Number
- 0001811755-2014-02580
- Event Type
- Malfunction
- Date Received
- July 21, 2014
- Date of Event
- June 23, 2014
- Report Date
- June 23, 2014
- Manufacturer
- STRYKER INSTRUMENTS-INAGI
- Product Code
- LFL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
A FOLLOW UP REPORT WILL BE FILED WHEN THE QUALITY INVESTIGATION HAS BEEN COMPLETED.
THE CUSTOMER COMMENT OF THE ANGLE JOINT BEING MISSING WAS CONFIRMED DURING SERVICE EVALUATION. DURING SERVICE EVALUATION IT WAS OBSERVED THE ANGLE NUT WAS MISSING DUE TO DAMAGE. A DAMAGED ANGLE NUT CAN BE CAUSED BY OVER TORQUING OF THE ANGLE NUT WHILE ATTACHING A TIP TO THE HANDPIECE.
IT WAS REPORTED THAT THE MALE THREADED TIP WAS MISSING WHEN THE 34KHZ SM LONG ANGLED W/O TIP WAS OPENED PRIOR TO A PROCEDURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY USING A BACK UP DEVICE. THERE WERE NO PATIENT OR USER INJURIES, AND NO ADVERSE CONSEQUENCES.
IT WAS REPORTED THAT THE MALE THREADED TIP WAS MISSING WHEN THE 34KHZ SM LONG ANGLED W/O TIP WAS OPENED PRIOR TO A PROCEDURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY USING A BACK UP DEVICE. THERE WERE NO PATIENT OR USER INJURIES, AND NO ADVERSE CONSEQUENCES.