FDA Adverse Event Injury Summary report: N

OT ULTRAMINI METER

MDR report key: 3946033 · Received July 21, 2014

Report

Report Number
2939301-2014-17990
Event Type
Injury
Date Received
July 21, 2014
Report Date
July 11, 2014
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K061118
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE METER INVOLVED WITH THIS COMPLAINT HAS BEEN RETURNED ON 07/25/2014 AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 07/29/2014 WITH THE FOLLOWING FINDINGS: THE METER HAS PASSED TESTING WITH NO FAULTS FOUND. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Additional Manufacturer Narrative · 1

LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

ON (B)(6) 2014, THE LAY-USER/PATIENT CONTACTED LIFESCAN (LFS) USA, ALLEGING THAT SHE TESTING IN THE MEMORY MODE, USE ERROR ISSUE. THIS COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER CARE ADVOCATE (CA) DOCUMENTATION. THE REPORTED ISSUE BEGAN ON (B)(6) 2014 AT 9 AM. THE PATIENT CLAIMED THAT DUE TO THE TESTING IN THE MEMORY MODE ISSUE, SHE DEVELOPED SYMPTOMS DESCRIBED AS ¿SHAKY.¿ THE PATIENT DID NOT BASE HER DIABETES MANAGEMENT ON THE SUBJECT METER AND DID NOT REQUIRE ANY MEDICAL INTERVENTION. DURING TROUBLESHOOTING, THE ISSUE WAS RESOLVED WITH TRAINING. THIS COMPLAINT IS BEING REPORTED DUE TO A POSSIBLE USE ERROR AND BECAUSE THE PATIENT HAD SYMPTOM THAT CAN BE SUGGESTIVE OF HYPOGLYCEMIA AFTER SHE ENCOUNTERED THE TESTING IN MEMORY MODE ISSUE. THE LFS PRODUCTS WERE REPLACED AND REQUESTED BACK FOR INVESTIGATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
425318 OT ULTRAMINI METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3640810

Patients

Seq Age Sex Outcome Treatment
1 66 YR Life Threatening