FDA Adverse Event Malfunction Summary report: N

530G INSULIN PUMP

MDR report key: 3945911 · Received July 21, 2014

Report

Report Number
3004209178-2014-88004
Event Type
Malfunction
Date Received
July 21, 2014
Date of Event
June 19, 2014
Report Date
June 19, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P120010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ID
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS RECEIVED WITH NORMAL OPERATING CURRENTS AND NO UNEXPECTED OFF NO POWER, LOW BATTERY OR FAILED BATTERY TEST ALARMS NOTED. THE UNIT HAD CRACKED RESERVOIR TUBE LIP.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT SHE HAS CHANGED THE BATTERY ON HER INSULIN PUMP SEVERAL TIMES AND KEEPS GETTING A FAILED BATTERY ALARM. CUSTOMER'S CURRENT BLOOD GLUCOSE READING IS 446 MG/DL AND HAS TREATED WITH A MANUAL INJECTION. NO FURTHER INFORMATION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
425925 530G INSULIN PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-751NAH

Patients

Seq Age Sex Outcome Treatment
1 29 YR