- Report Number
- 2939301-2014-17862
- Event Type
- Malfunction
- Date Received
- July 21, 2014
- Report Date
- July 10, 2014
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K082590
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP # 1 ¿ (10/03/2014). THE LAY USER/PATIENT¿S METER HAS BEEN RETURNED ON 08/29/2014 AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 09/30/2014 WITH THE FOLLOWING FINDINGS:THE METER PASSED ALL TESTING WITH NO FAULTS FOUND. THE REPORTED ISSUE COULD NOT BE REPRODUCED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
ON (B)(6) 2014, THE REPORTER CONTACTED LIFESCAN USA ALLEGING THAT THE SUBJECT METER (ONE TOUCH PING) READ INACCURATELY HIGH COMPARED TO ANOTHER DEVICE (ACCU-CHEK NANO). THE REPORTER CLAIMED OBTAINING A BLOOD GLUCOSE READING OF 500MG/DL WITH THE SUBJECT METER AND 290MG/DL ON THE OTHER DEVICE, PERFORMED WITHIN 30 MINUTES OF EACH OTHER. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE OF THESE GLUCOSE RESULTS EXCEEDS LIFESCAN¿S CRITERIA FOR ACCURACY. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 424905 | OT PING METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 3599246 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |