S4-ETL
Report
- Report Number
- 3006695864-2014-00350
- Event Type
- Injury
- Date Received
- July 21, 2014
- Date of Event
- December 26, 2013
- Report Date
- June 16, 2014
- Manufacturer
- ABBOTT MEDICAL OPTICS
- Product Code
- LZS
- PMA / PMN Number
- P930016
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- PHYSICIAN
Narratives
DATE OF EVENT WAS (B)(6) 2013. PLACEHOLDER.
(B)(4). MANUFACTURING DATE NOT AVAILABLE AT THE TIME OF THIS REPORT. (B)(4) 2014, APPLICATION SUPPORT MANAGER VISITED ACCOUNT AND OBSERVED SURGEON COMPLETE 1 ILASIK PATIENT (BOTH EYES ) SUCCESSFULLY AND DID NOT SEE ANYTHING TO LEAD TO UNDERCORRECTIONS. CONFIRMED INNER DIAMETER OF SPHERE LESS THAN 0.80. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT MEDICAL OPTICS HAS BEEN SUBMITTED. PLACEHOLDER.
ACCOUNT REPORTED RECEIVED SURGEONS HAVE BEEN NOTICING AN INCREASE IN UNDERCORRECTIONS THE LAST FEW MONTHS. WE HAVE CONSIDERED ALL THE POSSIBLE VARIABLES THAT MIGHT ACCOUNT FOR THEM BUT THEY ARE DIFFERENT SURGEONS, DIFFERENT DAYS, DIFFERENT TEAMS, DIFFERENT CORRECTIONS. PATIENT REQUIRED AN ENHANCEMENT PROCEDURE FOR THE CORRECTION OF THE UNDERCORRECTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 425086 | S4-ETL | EXCIMER LASER | LZS | ABBOTT MEDICAL OPTICS | 0030-4077 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 28 YR | Required Intervention |