FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3945556 · Received July 21, 2014

Report

Report Number
2032227-2014-04530
Event Type
Malfunction
Date Received
July 21, 2014
Date of Event
June 15, 2014
Report Date
June 17, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO MOISTURE DAMAGE ON KEYPAD TRACES. NO MOISTURE DAMAGE FOUND ON ELECTRONIC ASSEMBLY, LCD BOARD OR MOTOR. THE INSULIN PUMP HAD A CRACKED DISPLAY WINDOW CORNER, CRACKED BATTERY TUBE THREADS, CRACKED BELT CLIP SLOT AND CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

CUSTOMER REPORTED THE HE ACCIDENTLY JUMPED IN THE POOL AND THE BUTTONS ON THE PUMP ARE NOT WORKING. CUSTOMER'S BLOOD GLUCOSE WAS 190 MG/DL. THE ACT BUTTON ON THE INSULIN PUMP IS NOT RESPONDING. CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE DEVICE AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
425031 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723LNAS

Patients

Seq Age Sex Outcome Treatment
1 23 YR