FDA Adverse Event Malfunction Summary report: N

COULTER LH 750 HEMATOLOGY ANALYZER

MDR report key: 3945538 · Received July 21, 2014

Report

Report Number
1061932-2014-01666
Event Type
Malfunction
Date Received
July 21, 2014
Date of Event
June 28, 2014
Report Date
June 28, 2014
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K011342
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ON (B)(4) 2014, A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) EVALUATED THE ANALYZER AND FOUND DAMAGED TUBING AT THE NEEDLE BACK WASH THAT CAUSED THE LEAK. THE FSE REPLACED THE TUBING AND ASSOCIATED PINCH VALVES AND THE INSTRUMENT RAN WITHOUT LEAKS. THE REPAIRS WERE VERIFIED PER ESTABLISHED PROCEDURES. (B)(4).

Description of Event or Problem · 1

CUSTOMER REPORTED A LEAK WHEN USING THE COULTER LH 750 HEMATOLOGY ANALYZER. THE LEAK WAS DESCRIBED AS ABOUT 100 ML OF LIQUID THAT LEAKED UNDER THE INSTRUMENT. THE LEAK WAS NOT CONTAINED WITHIN THE INSTRUMENT. A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS SENT TO THE CUSTOMER'S FACILITY TO EVALUATE THE ANALYZER. THE OPERATOR WAS WEARING GLOVES, LAB COAT, AND EYEGLASSES AT THE TIME OF THE INCIDENT. THERE WAS NO BIOHAZARD EXPOSURE TO OPEN WOUNDS OR MUCOUS MEMBRANES. THERE WAS NO REPORT OF ERRONEOUS TEST RESULTS ASSOCIATED WITH THIS EVENT. THERE WAS NO DEATH, INJURY OR AFFECT TO USER OR PATIENT TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
424648 COULTER LH 750 HEMATOLOGY ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER NA NA

Patients

Seq Age Sex Outcome Treatment
1