FDA Adverse Event Malfunction Summary report: N

VERSION CONTROL STEM INSERTER

MDR report key: 3945414 · Received July 21, 2014

Report

Report Number
0002249697-2014-02802
Event Type
Malfunction
Date Received
July 21, 2014
Date of Event
June 25, 2014
Report Date
June 25, 2014
Manufacturer
STRYKER ORTHOPAEDICS-MAHWAH
Product Code
LXH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

CORRECTED DATA: DEVICE NOT RETURNED FOR EVALUATION A REVIEW OF THE DEVICE HISTORY AND COMPLAINT HISTORY RECORDS COULD NOT BE PERFORMED BECAUSE THE DEVICE ASSOCIATED WITH THIS EVENT WAS NOT RETURNED NOR WAS A VALID LOT CODE PROVIDED. VISUAL, DIMENSIONAL AND FUNCTIONAL ANALYSIS COULD NOT BE PERFORMED AS THE DEVICE WAS NOT RETURNED. THE EVENT COULD NOT BE CONFIRMED NOR THE ROOT CAUSE OF THE REPORTED EVENT DETERMINED DUE TO THE MINIMAL INFORMATION RECEIVED. NO FURTHER INVESTIGATION FOR THIS EVENT IS POSSIBLE AT THIS TIME AS NO DEVICES AND INSUFFICIENT INFORMATION WAS RECEIVED BY STRYKER ORTHOPAEDICS. IF DEVICES AND / OR ADDITIONAL INFORMATION BECOME AVAILABLE, THIS INVESTIGATION WILL BE REOPENED.

Additional Manufacturer Narrative · 1

WHEN COMPLETED, THE INVESTIGATION RESULTS WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

DURING THE INSERTION OF A PLASMA BOWED 167MM STEM THE VERSION CONTROL HANDLE WOULD NOT DISENGAGE THE STEM. SUBSEQUENTLY THE INTERNAL SHAFT OF THE THREADED INSERTER BROKE. THE THREADED PORTION WAS CONNECTED TO THE STEM WHILE THE HANDLE WAS REMOVED. SURGEON USED VICE GRIPS TO REMOVE THE THREADED SHAFT.

Description of Event or Problem · 1

DURING THE INSERTION OF A PLASMA BOWED 167MM STEM THE VERSION CONTROL HANDLE WOULD NOT DISENGAGE THE STEM. SUBSEQUENTLY THE INTERNAL SHAFT OF THE THREADED INSERTER BROKE. THE THREADED PORTION WAS CONNECTED TO THE STEM WHILE THE HANDLE WAS REMOVED. SURGEON USED VICE GRIPS TO REMOVE THE THREADED SHAFT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
425570 VERSION CONTROL STEM INSERTER INSTRUMENT LXH STRYKER ORTHOPAEDICS-MAHWAH UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 85 YR Other