FDA Adverse Event Malfunction Summary report: N

CARE ASSIST BED

MDR report key: 3945250 · Received May 6, 2014

Report

Report Number
9615739-2014-00188
Event Type
Malfunction
Date Received
May 6, 2014
Date of Event
April 8, 2014
Report Date
April 8, 2014
Manufacturer
HILL-ROM PLUVIGNER
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM TECHNICIAN FOUND THE RIGHT HEAD BRAKE NOT HOLDING MOST LIKELY DUE TO NORMAL WEAR AND TEAR. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2014. IT IS UNK IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN REPLACED THE RIGHT HEAD CASTER AND SUPPORT TO RESOLVE THE ISSUE. BASED ON THIS INFO, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM TECHNICAL SERVICE RECEIVED A REPORT FROM THE ACCOUNT STATING THE BRAKES ARE NOT HOLDING. THE BED WAS LOCATED AT THE ACCOUNT IN MAINTENANCE. THERE WAS NO PT/USER INJURY REPORTED. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
271011 CARE ASSIST BED A/C POWERED ADJUSTABLE HOSP BED FNL HILL-ROM PLUVIGNER 1170