DIMENSION® CLINICAL CHEMISTRY SYSTEM
Report
- Report Number
- 2517506-2014-00065
- Event Type
- Malfunction
- Date Received
- July 18, 2014
- Date of Event
- June 29, 2014
- Report Date
- June 29, 2014
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
- Product Code
- MMI
- PMA / PMN Number
- K963498
- Removal / Correction Number
- 2517506-07/15/2014-003-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
SIEMENS HEALTHCARE DIAGNOSTICS HAS CONFIRMED COMPLAINTS OF QC AND PATIENT SAMPLE RESULT ACCURACY SHIFTS ON TROPONIN I CTNI (RF421C) AND LTNI (RF521C); FREET4 (RF410); TSH (RF412); AND N-TERMINAL PRO-BRAIN NATRIURETIC PEPTIDE PBNP (RF423A) AND LPBN (RF523A) WHEN USING DIMENSION® CHEMISTRY WASH LOT RD41041. INTERNAL INVESTIGATION HAS CONFIRMED BACTERIAL CONTAMINATION IN CHEMISTRY WASH LOT RD41041. A POSITIVE BIAS WILL BE OBSERVED ON CTNI, LTNI, TSH, PBNP AND LPBN AND A NEGATIVE BIAS WILL BE SEEN ON FT4 WHEN THIS LOT IS FIRST PLACED ON THE INSTRUMENT WITHOUT RECALIBRATION OF METHODS. RECALIBRATION OF THESE ASSAYS WHEN USING THIS LOT WILL NOT CORRECT THE PROBLEM AND MAY CAUSE FALSELY HIGH OR LOW RESULTS. SIEMENS ISSUED AN URGENT FIELD SAFETY NOTICE DATED JULY 2014 TO ALL ACCOUNTS WHO HAD RECEIVED TO THE IMPACTED LOT. CUSTOMERS WERE ADVISED OF THE IMMEDIATE REMEDIAL ACTIONS. IMMEDIATE ACTIONS TO BE UNDERTAKEN BY THE CUSTOMER: IMMEDIATELY DISCONTINUE USE AND DISCARD ALL INVENTORY OF CHEMISTRY WASH LOT RD41041. - IF THIS LOT WAS NEVER IN USE ON THE INSTRUMENT, NO FURTHER ACTION IS NEEDED. - IF THIS LOT HAS BEEN IN USE, CONTINUE WITH STEPS 2 THROUGH 5. PLACE AN ALTERNATE LOT OF CHEMISTRY WASH ON THE DIMENSION SYSTEM. PRIME THE NEW BOTTLE OF CHEMISTRY WASH 30 TIMES. RECALIBRATE THE CTNI, LTNI, TSH, FT4, PBNP AND/OR LPBN ASSAYS USING THE ALTERNATE LOT OF CHEMISTRY WASH ONBOARD. FOR THE LISTED ASSAYS, PROCESS QUALITY CONTROL EVERY 8 HOURS. SIEMENS IS SHIPPING REPLACEMENT PRODUCT AT NO CHARGE FOR ALL ORDERS FILLED WITH LOT RD41041. AS AN ADDITIONAL PRECAUTION, A SIEMENS CUSTOMER SERVICE ENGINEER WILL BE MAKING A VISIT TO DECONTAMINATE THE SYSTEM(S) OF CUSTOMERS WHO HAVE USED LOT RD41041.
FALSELY ELEVATED TROPONIN I (CTNI), THYROID STIMULATING HORMONE (TSH); AND N-TERMINAL PRO-BRAIN NATRIURETIC PEPTIDE (PBNP) RESULTS WERE OBTAINED ON QC SAMPLES WITH CHEMISTRY WASH LOT RD41041 ON THE INSTRUMENT. NO PATIENT RESULTS WERE REPORTED TO PHYSICIANS. LOWER RESULTS WITHIN LABORATORY EXPECTATIONS WERE OBTAINED WITH AN ALTERNATE LOT OF CHEMISTRY WASH. PATIENT TREATMENT WAS NOT ALTERED OR PRESCRIBED ON THE BASIS OF THE FALSELY ELEVATED QC RESULTS. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED QC RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 422749 | DIMENSION® CLINICAL CHEMISTRY SYSTEM | CHEM WASH | MMI | SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW | RD41041 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |