PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-04605
- Event Type
- Malfunction
- Date Received
- July 18, 2014
- Date of Event
- June 18, 2014
- Report Date
- June 18, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA
- Reporter Occupation
- PATIENT
Narratives
INSULIN PUMP RECEIVED WITH BUTTON ERROR ALARM AND NO BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. INSULIN PUMP RECEIVED WITH MINOR SCRATCHED DISPLAY WINDOW, CRACKED BELT CLIP SLOT, CRACKED RESERVOIR TUBE LIP AND MISSING END CAP STICKER.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
THE PATIENT STATED THAT THE PUMP ALARMED BUTTON ERROR, AND DELIVERED TOO MUCH INSULIN. THE PATIENT STATED THAT SHE WAS PRIMING THE PUMP, AND NOTICED THAT THE NUMBER OF UNITS ON THE SCREEN REACHED 29. THE PATIENT TOOK THE PUMP OFF, AND THEN IT GAVE A BUTTON ERROR ALARM, AND THE ACT BUTTON STOPPED RESPONDING. THE REPORTED BG AT THE TIME OF THE CALL WAS 230 MG/DL. ADVISED THE PATIENT THAT THE PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 422634 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-722LNAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR |