PINN MAR LIP LNR 32IDX54OD
Report
- Report Number
- 1818910-2014-23793
- Event Type
- Injury
- Date Received
- July 18, 2014
- Date of Event
- June 26, 2014
- Report Date
- July 22, 2014
- Manufacturer
- DEPUY ORTHOPAEDICS, INC.1818910
- Product Code
- LPH
- PMA / PMN Number
- PK000306
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
Narratives
: THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. A COMPLAINT DATABASE SEARCH FINDS NO OTHER REPORTED INCIDENTS AGAINST THE PROVIDED PRODUCT AND LOT COMBINATIONS. REQUESTS FOR ADDITIONAL INVESTIGATIONAL INPUTS WERE MADE IN ACCORDANCE WITH WI-7915 APPENDIX A. NO ADDITIONAL INFORMATION WAS OBTAINED. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO IDENTIFY ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
THE PATIENT WAS REVISED BECAUSE OF DISASSOCIATION.
PATIENT HAD LEFT TOTAL HIP ARTHROPLASTY REVISION DUE TO LINER DISASSOCIATION FROM ACETABULAR COMPONENT. IT ALSO INDICATES BLACK STAINED TISSUE C/W METALLOSIS. DATE OF IMPLANT HAS BEEN PROVIDED AND AN INVOICE LOCATED. PART AND LOT NUMBERS ARE BEING UPDATED AND A FEMORAL HEAD IS NOW BEING REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 421286 | PINN MAR LIP LNR 32IDX54OD | HIP ACETABULAR CUP | LPH | DEPUY ORTHOPAEDICS, INC.1818910 | CD8FG1000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 79 YR | Required Intervention |