FDA Adverse Event Malfunction Summary report: N

G4 PLATINUM PEDIATRIC CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 3945098 · Received July 18, 2014

Report

Report Number
3004753838-2014-25017
Event Type
Malfunction
Date Received
July 18, 2014
Date of Event
June 21, 2014
Report Date
June 22, 2014
Manufacturer
DEXCOM, INC.
Product Code
MDS
PMA / PMN Number
P120005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION OF THE RETURNED DEVICE CONFIRMED THE REPORTED NO AUDIO OUTPUT. THE ROOT CAUSE WAS DETERMINED TO BE A DEFECTIVE G4 SPEAKER ASSEMBLY ASY000016.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

PATIENT CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2014 TO CLAIM NO AUDIO OUTPUT ON (B)(6) 2014. AT THE TIME OF THE CALL TO DEXCOM TECHNICAL SUPPORT, NO MEDICAL INTERVENTION OR INJURIES WERE REPORTED. ADDITIONALLY, DURING THE CALL, DEXCOM TECHNICAL SUPPORT ADVISED PATIENT TO TEST THE ALERT FUNCTIONALITY AND REPORTED ALERTS WERE NOT FUNCTIONAL, BUT DEVICE DID VIBRATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
421285 G4 PLATINUM PEDIATRIC CONTINUOUS GLUCOSE MONITORING SYSTEM MDS MDS DEXCOM, INC. MT22430

Patients

Seq Age Sex Outcome Treatment
1 15 YR