FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3945093 · Received July 18, 2014

Report

Report Number
2032227-2014-04514
Event Type
Malfunction
Date Received
July 18, 2014
Date of Event
June 18, 2014
Report Date
June 18, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED THE PRIME, DISPLACEMENT, BASIC OCCLUSION, OCCLUSION, AND EXCESSIVE NO DELIVERY ALARM TEST. NO MOTOR ERROR ALARMS WERE NOTED. THE MOTOR TESTED OKAY. A CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, MINOR SCRATCHES ON THE DISPLAY WINDOW AND A CRACKED CASE NEAR DISPLAY WINDOW CORNERS WERE NOTED DURING A VISUAL INSPECTION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER RECEIVED A MOTOR ERROR ALARM ON HER INSULIN PUMP. THE CUSTOMER WAS PROVIDED ASSISTANCE WITH CLEARING THE ALARM, ADVISED TO DISCONNECT FROM THE INSULIN PUMP AND WAS UNABLE TO REWIND IT. HER BLOOD GLUCOSE LEVEL WAS 87 MG/DL. ADVICE WAS GIVEN TO THE CUSTOMER TO DISCONTINUE PUMP USAGE AND REVERT TO A BACK-UP PLAN. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
421882 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-523RNAP

Patients

Seq Age Sex Outcome Treatment
1 62 YR