FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3945056 · Received July 18, 2014

Report

Report Number
3004209178-2014-87786
Event Type
Malfunction
Date Received
July 18, 2014
Date of Event
June 18, 2014
Report Date
June 18, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

INSULIN PUMP ALARMED A MOTION SENSOR TEST FAILURE DURING REWIND, DUE TO OUT OF PHASE MOTOR ENCODER SIGNAL. UNABLE TO PERFORM DISPLACEMENT TEST AND CONFIRM MOTOR ERROR DUE TO ALARM. INSULIN PUMP RECEIVED WITH CRACKED RESERVOIR TUBE LIP.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A MOTOR ERROR ALARM. THE BLOOD GLUCOSE READING WAS 220 MG/DL. THE PATIENT DOES NOT RECALL ANY SIGNIFICANT EVENTS LEADING TO THE ALARM. THE INSULIN PUMP WAS NOT EXPOSED TO A STRONG MAGNETIC FIELD OR AN MRI. THE CUSTOMER WAS ASSISTED WITH CLEARING THE ALARM. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO A BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
421883 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523LNAP

Patients

Seq Age Sex Outcome Treatment
1 75 YR