FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3945042 · Received July 18, 2014

Report

Report Number
3004209178-2014-87797
Event Type
Malfunction
Date Received
July 18, 2014
Date of Event
June 16, 2014
Report Date
June 18, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP RECEIVED WITHOUT ORIGINAL BATTERY CAP AND UNABLE TO VERIFY CRACKED BATTERY CAP. THE INSULIN PUMP HAD A BROKEN BATTERY TUBE THREADS, CRACKED RESERVOIR TUBE LIP AND MISSING END CAP STICKER.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP WAS DAMAGED, REQUIRING THE CUSTOMER TO TAPE THE BATTERY CAP DOWN. THE BLOOD GLUCOSE READING WAS 49 MG/DL. THE CUSTOMER'S MOTHER STATED THAT THE DEVICE AND THE BATTERY CAP WERE CRACKED FROM NORMAL WEAR AND TEAR. ADVISED DISCONTINUATION AND REPLACEMENT OF THE INSULIN PUMP. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
421755 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 20 YR