FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3944760 · Received July 18, 2014

Report

Report Number
2531779-2014-20706
Event Type
Malfunction
Date Received
July 18, 2014
Report Date
July 12, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/27/2014 WITH THE FOLLOWING FINDINGS: DURING TESTING, THE PUMP POWERED ON TO A FUNCTIONING DISPLAY SCREEN. THERE WAS EVIDENCE OF MOISTURE BEHIND THE DISPLAY SCREEN. UNRELATED TO THE INITIAL COMPLAINT, THE BATTERY COMPARTMENT WAS CRACKED. THE LEAK TEST FAILED DUE TO A BATTERY COMPARTMENT LEAK. THE PUMP WAS OPENED AND THERE WAS EVIDENCE OF MOISTURE ON INTERNAL COMPONENTS OF THE PUMP.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING THAT THE DISPLAY SCREEN WAS BLANK WITH EVIDENCE OF MOISTURE BEHIND THE DISPLAY. THERE WAS NO INDICATION THAT THE DEVICE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
422322 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 15 YR