ACUSON
Report
- Report Number
- 3009498591-2014-00014
- Event Type
- Malfunction
- Date Received
- May 5, 2014
- Date of Event
- February 10, 2014
- Report Date
- May 2, 2014
- Manufacturer
- SIEMENS MEDICAL SOLUTIONS USA INC.
- Product Code
- IYN
- PMA / PMN Number
- 123622
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THERE WAS A CODING ERROR INTRODUCED WITH A CHANGE TO SUPPORT BIPLANE IMAGING FOR THE 4.0 RELEASE (BEFORE THE 3.5 RELEASE WAS BRANCHED FROM THE 4.0 RELEASE). THE SCOPE OF A PARAMETER TO HANDLE CHANGING MULTIHZ IN DOPPLER WAS CHANGED WITHOUT A CORRESPONDING CHANGE IN THE CODE WHICH DETERMINES THAT THE STRIP SHOULD BE RESET. SO THE STRIP WAS NOT RESET AS IT SHOULD HAVE BEEN AND THE DOPPLER DATA FROM THE OLD FREQUENCY IS STILL DISPLAYED IN THE DOPPLER STRIP. THE LIKELIHOOD OF THE USER CHANGING THE DOPPLER FREQUENCY DURING SPECTRAL DOPPLER, THEN FREEZING WHILE THE OLD CINE STRIP DATA IS DISPLAYED (ON THE RIGHT-HAND SIDE) IS LOW. THIS WORKFLOW IS A RARE AND CONSIDERED A CORNER CASE.
DURING IN-HOUSE TESTING, DOPPLER FREQUENCY CHANGE DOES NOT RESTART THE STRIP, BUT RATHER ALLOWS TWO DIFFERENT RESULTS BASED ON SCALE CHANGES. WHEN YOU CHANGE FREQUENCY DURING LIVE SPECTRAL DISPLAY, THE SPECTRAL STRIP SHOULD REFRESH, STARTING A NEW BLANK STRIP. INSTEAD THE OLD STRIP INFO IN RETAINED ON THE SCREEN. YOU SEE BOTH THE OLD DOPPLER DATA FROM THE PREVIOUS FREQUENCY DISPLAYED ON THE SAME STRIP AS THE NEW DOPPLER DATA WITH A BLACK LINE DIVIDING THEM. THE SCALE CHANGES AND IS CORRECT FOR THE NEW DOPPLER DATE. IN CINE, IF YOU MEASURE THE SAME THE DOPPLER DATA (E.G. PEAK VELOCITY) FROM BOTH THE OLD AND NEW DOPPLER FREQUENCIES, YOU WILL RECEIVE TWO DIFFERENT VALUES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 268058 | ACUSON | DIAGNOSTIC ULTRASOUND DEVICE | IYN | SIEMENS MEDICAL SOLUTIONS USA INC. | ACUSON SC2000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |