INFUSOR
Report
- Report Number
- 1416980-2014-23276
- Event Type
- Malfunction
- Date Received
- July 18, 2014
- Date of Event
- June 24, 2014
- Report Date
- June 24, 2014
- Manufacturer
- BAXTER HEALTHCARE - IRVINE
- Product Code
- MEB
- PMA / PMN Number
- K062457
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
(B)(4). A CAPA HAS BEEN INITIATED TO INVESTIGATE THIS ISSUE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
(B)(4). IF ADDITIONAL RELEVANT INFORMATION IS OBTAINED, THEN A FOLLOW-UP MDR WILL BE SUBMITTED.
(B)(4). THE LOT WAS MANUFACTURED BETWEEN FEBRUARY 14, 2014 AND FEBRUARY 17, 2014. EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. VISUAL INSPECTION NOTED SOLID, WHITE PARTICLES, BETWEEN 0.20 AND 0.45 MM IN SIZE, FLOATING IN THE FLUID OF THE RESERVOIR. FOURIER TRANSFORM INFRARED SPECTROSCOPY (FT-IR) SCANNING IDENTIFIED THE PARTICLES TO BE POLYESTER MATERIAL. A REVIEW OF ALL BATCH RECORD DOCUMENTS WAS PERFORMED WITH NO ISSUES NOTED DURING THE MANUFACTURING PROCESS. THERE WERE NO DEVIATIONS FROM STANDARD PROCEDURE AND NO EXCEPTIONS RELATED TO THE REPORTED CONDITION WERE NOTED. THE CAUSE OR ORIGIN OF THE PARTICLES WAS UNABLE TO BE DETERMINED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT AN INTERMATE HAD PARTICULATE MATTER INSIDE ITS BALLOON. THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS REQUESTED AND IS NOT AVAILABLE.