ACUSON
Report
- Report Number
- 3009498591-2014-00019
- Event Type
- Malfunction
- Date Received
- May 9, 2014
- Date of Event
- April 17, 2014
- Report Date
- May 8, 2014
- Manufacturer
- SIEMENS MEDICAL SOLUTIONS USA INC.
- Product Code
- IYN
- PMA / PMN Number
- 123622
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THIS ISSUE WAS INTRODUCED WHEN WE FIXED US_00113852 (VC30A) / US_00114716 (VC25D); "MEASUREMENT VALUE CHANGES IN THE REPORT WHEN BASELINE IS CHANGED". WE REMOVED THE DEPENDENCE ON SOME PARAMETER UPDATES IN ORDER TO FIX THOSE ISSUES.
WHILE SCANNING A CAROTID EXAM, USER PLACED A PULSED WAVEGATE WITHIN A STENOSIS IN THE EXTERNAL CAROTID ARTERY. SHE INCREASED THE PW PRF TO 13,021 (SLIGHTLY ABOVE 300 CM/S). THE AUTO-STAT MEASUREMENT CALIPER IS APPROPRIATELY PLACED (NEAR 300 CM/S) BUT THE DISPLAYED MEASUREMENT OF THE PEAK SYSTOLE VELOCITY IS 101.5 CM/S. THIS ISSUE IS REPRODUCTIBLE IN ALL EXAMS (EXCEPT CARDIAC) BY CHANGING MULTIHZ OR BASELINE OR SCALE WITH AUTO-STATS ENABLED. ON FREEZE, THE VELOCITY WILL DISPLAY FROM THE PREVIOUS SCALE. ALL VALUES IN THE MDA WILL BE INCORRECT, ANOTHER SCENARIO IS TO USE VELOCITY TOOL ON A LIVE IMAGE AND CHANGE MULTIHZ, THE VALUE DOES NT UPDATE IN THIS CASE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 281865 | ACUSON | DIAGNOSTIC ULTRASOUND DEVICE | IYN | SIEMENS MEDICAL SOLUTIONS USA INC. | ACUSON SC2000 |