FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 3944528
·
Received May 9, 2014
Report
- Report Number
- 3008642652-2014-01400
- Event Type
- Malfunction
- Date Received
- May 9, 2014
- Date of Event
- April 14, 2014
- Report Date
- May 6, 2014
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. UPON EVALUATION, THE CABLE CONNECTING THE DISTRIBUTION NODE TO THE REAR THERAPY ELECTRODES WAS PULLED FROM THE STRAIN RELIEF, DAMAGING WIRES. THE ROOT CAUSE FOR THE STRAINED CABLE COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE ELECTRODE BELT. THE LAST PATIENT TO USE THIS ELECTRODE BELT DID NOT REPORT ANY DEFICIENCIES.
Description of Event or Problem · 1
A REVIEW OF SERVICE DATA DETECTED A REPORTABLE PROBLEM. DURING SERVICING, ELECTRODE BELT SN (B)(4) HAD DAMAGED CABLES AND FAILED INCOMING FUNCTIONALITY TESTING. THE LAST PATIENT TO USE THIS ELECTRODE BELT DID NOT REPORT ANY DEFICIENCIES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 281185 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |