FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 3944507
·
Received May 9, 2014
Report
- Report Number
- 3008642652-2014-01397
- Event Type
- Malfunction
- Date Received
- May 9, 2014
- Date of Event
- April 17, 2014
- Report Date
- May 6, 2014
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF BATTERY CHARGER/MODEM SN (B)(4) HAS BEEN COMPLETED. UPON EVALUATION THE BATTERY CHARGER/MODEM WAS UNABLE TO POWER UP. THE CAUSE FOR THE FAILURE WAS ISOLATED TO A DEFECTIVE POWER SUPPLY BRICK. THE ROOT CAUSE FOR THE DEFECTIVE POWER SUPPLY BRICK COULD NOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BATTERY CHARGER/MODEM. THE LAST PATIENT TO USE THIS BATTERY CHARGER/MODEM DID NOT REPORT ANY DEFICIENCIES.
Description of Event or Problem · 1
A REVIEW OF SERVICE DATA DETECTED A REPORTABLE PROBLEM. DURING SERVICING, BATTERY CHARGER/MODEM WAS UNABLE TO POWER UP. THE LAST PATIENT TO USE THIS BATTERY CHARGER/MODEM DID NOT REPORT ANY DEFICIENCIES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 281000 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 |