LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3008642652-2014-01407
- Event Type
- Malfunction
- Date Received
- May 9, 2014
- Date of Event
- April 17, 2014
- Report Date
- May 6, 2014
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. UPON EVALUATION THE ELECTRODE BELT FAILED A THERAPY ELECTRODE RECOGNITION TEST. THE CAUSE FOR THE FAILURE WAS ISOLATED TO AN OPEN PULSE WIRE IN THE CABLE CONNECTING ECG ''A'' AND ECG ''B''. THERE WAS NO SIGN OF EXTERNAL DAMAGE TO THE CABLE. THE ROOT CAUSE FOR THE OPEN PULSE WIRE COULD NOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE ELECTRODE BELT. THE LAST PATIENT TO USE THIS ELECTRODE BELT DID NOT REPORT ANY DEFICIENCIES.
A REVIEW OF SERVICE DATA DETECTED A REPORTABLE PROBLEM. DURING SERVICING, ELECTRODE BELT SN (B)(4) FAILED INCOMING FUNCTIONALITY TESTING. THE LAST PATIENT TO USE THIS ELECTRODE BELT DID NOT REPORT ANY DEFICIENCIES
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 281182 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |