FDA Adverse Event Injury Summary report: N

ARTICULEZE M HEAD 36MM +5

MDR report key: 3944428 · Received July 18, 2014

Report

Report Number
1818910-2014-23764
Event Type
Injury
Date Received
July 18, 2014
Date of Event
June 18, 2014
Report Date
June 19, 2014
Manufacturer
DEPUY ORTHOPAEDICS INC US
Product Code
JDI
PMA / PMN Number
PK980513
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL INFO: THIS COMPLAINT IS STILL CONSIDERED CLOSED AT THIS TIME. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE: THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE: THE DEVICE ASSOCIATED WITH THIS REPORT WAS NOT RETURNED. A COMPLAINT DATABASE SEARCH FINDS NO OTHER REPORTED INCIDENTS AGAINST THE PROVIDED PRODUCT AND LOT COMBINATION. REQUESTS FOR ADDITIONAL INVESTIGATIONAL INPUTS WERE MADE IN ACCORDANCE WITH WI-7915 APPENDIX A. NO ADDITIONAL INFORMATION WAS OBTAINED. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO IDENTIFY ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Additional Manufacturer Narrative · 1

THE INVESTIGATION REMAINS CLOSED, AS THE ADDED INFORMATION DOES NOT CHANGE THE OUTCOME OF THE INVESTIGATION.

Additional Manufacturer Narrative · 1

UDI: (B)(4).

Description of Event or Problem · 1

THE PATIENT WAS REVISED BECAUSE OF METAL WEAR.

Description of Event or Problem · 1

UPDATE 6/29/15-PFS AND MEDICAL RECORDS RECEIVED. AFTER REVIEW OF THE MEDICAL RECORDS FOR MDR REPORTABILITY, THE REVISION OPERATIVE NOTE INDICATED PAIN, INCREASED METAL IONS AND METALLOSIS/METAL DEBRIS. LAB RESULTS FROM (B)(6) 2014 CONFIRMED THE HIGH METAL IONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
421446 ARTICULEZE M HEAD 36MM +5 HIP FEMORAL HEAD JDI DEPUY ORTHOPAEDICS INC US 2333229

Patients

Seq Age Sex Outcome Treatment
1 78 YR Required Intervention