FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 3944169
·
Received May 7, 2014
Report
- Report Number
- 3008642652-2014-01391
- Event Type
- Malfunction
- Date Received
- May 7, 2014
- Date of Event
- April 7, 2014
- Report Date
- May 6, 2014
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (ADD GEL/REPLACE BELT MESSAGES) WAS CONFIRMED. UPON INVESTIGATION, THE DISTRIBUTION NODE (DN) TO REAR THERAPY ELECTRODE (TE) CABLE WAS PULLED FROM STRAIN RELIEF. THE CAUSE FOR THE ALARMS WAS THE DAMAGED CABLES. THE ROOT CAUSE OF THE DAMAGED CABLES COULD NOT BE POSITIVELY IDENTIFIED, BUT WAS LIKELY EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE ELECTRODE BELT. THE PATIENT WAS ISSUED A REPLACEMENT ELECTRODE BELT.
Description of Event or Problem · 1
A (B)(6) YEAR OLD MALE PATIENT'S NURSE CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT ADD GEL/REPLACE BELT MESSAGES. THE PATIENT WAS ISSUED A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 274938 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 71 YR |