FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 3000 SYSTEM

MDR report key: 3944160 · Received May 7, 2014

Report

Report Number
3008642652-2014-01388
Event Type
Malfunction
Date Received
May 7, 2014
Date of Event
April 7, 2014
Report Date
May 6, 2014
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION OF MONITOR SN (B)(4) HAS BEEN COMPLETED. UPON EVALUATION THE MONITOR FAILED A PULSE TEST. THE CAUSE FOR THE FAILURE WAS ISOLATED TO CORROSION ON THE MONITOR DEFIBRILLATOR BOARD. THE ROOT CAUSE FOR THE CORROSION WAS INGRESS OF AN UNK LIQUID. NO ADVERSE EVENT RESULTED FROM THE CORROSION. THE LAST PATIENT TO USE THIS MONITOR DID NOT REPORT ANY DEFICIENCIES.

Description of Event or Problem · 1

A REVIEW OF SERVICE DATA DETECTED A REPORTABLE PROBLEM. DURING SERVICING, MONITOR SN (B)(4) FAILED A PULSE TEST. THE LAST PATIENT TO USE THIS MONITOR DID NOT REPORT ANY DEFICIENCIES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
274844 LIFEVEST WCD 3000 SYSTEM MVK ZOLL LIFECOR CORPORATION WCD 3000 NA

Patients

Seq Age Sex Outcome Treatment
1 NA