FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 3000 SYSTEM
MDR report key: 3944160
·
Received May 7, 2014
Report
- Report Number
- 3008642652-2014-01388
- Event Type
- Malfunction
- Date Received
- May 7, 2014
- Date of Event
- April 7, 2014
- Report Date
- May 6, 2014
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION OF MONITOR SN (B)(4) HAS BEEN COMPLETED. UPON EVALUATION THE MONITOR FAILED A PULSE TEST. THE CAUSE FOR THE FAILURE WAS ISOLATED TO CORROSION ON THE MONITOR DEFIBRILLATOR BOARD. THE ROOT CAUSE FOR THE CORROSION WAS INGRESS OF AN UNK LIQUID. NO ADVERSE EVENT RESULTED FROM THE CORROSION. THE LAST PATIENT TO USE THIS MONITOR DID NOT REPORT ANY DEFICIENCIES.
Description of Event or Problem · 1
A REVIEW OF SERVICE DATA DETECTED A REPORTABLE PROBLEM. DURING SERVICING, MONITOR SN (B)(4) FAILED A PULSE TEST. THE LAST PATIENT TO USE THIS MONITOR DID NOT REPORT ANY DEFICIENCIES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 274844 | LIFEVEST WCD 3000 SYSTEM | MVK | ZOLL LIFECOR CORPORATION | WCD 3000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |