HERCULES 3 STAGE BALLOON ESOPHAGEAL
Report
- Report Number
- 1037905-2014-00161
- Event Type
- Malfunction
- Date Received
- May 1, 2014
- Report Date
- April 3, 2014
- Manufacturer
- WILSON-COOK MEDICAL INC.
- Product Code
- KNQ
- PMA / PMN Number
- K061937
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- NURSE
Narratives
INVESTIGATION EVAL: OUR EVAL OF THE RETURNED DEVICE CONFIRMED THE REPORT. THE BALLOON WAS INFLATED WITH WATER USING A 60CC SYRINGE AND AN INFLATION HANDLE. AS THE BALLOON WAS BEING INFLATED A STREAM OF WATER COULD BE SEEN EXITING THE BALLOON. A SMALL PINHOLE WAS NOTICED NEAR THE CENTER OF THE BALLOON. NO PART OF THE BALLOON WAS MISSING. A PRODUCT SPECIFIC DISCREPANCY THAT COULD HAVE CAUSED OR CONTRIBUTED TO THIS OBSERVATION WAS NOT OBSERVED DURING OUR LAB ANALYSIS. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSIONS: A DEFINITIVE CAUSE FOR THIS OBSERVATION COULD NOT BE DETERMINED BECAUSE THE ACTUAL USE CONDITIONS COULD NOT BE DUPLICATED IN THE LAB SETTING. DUE TO VARIETY OF CLINICAL CONDITIONS SUCH AS PT ANATOMY, ENDOSCOPE POSITION OR PROGRESSION OF DISEASE STATE, WE COULD NOT REPRODUCE THE ACTUAL CONDITIONS OF PRODUCT USAGE DURING OUR LAB ANALYSIS. THIS LIMITS OUR ABILITY TO CONCLUSIVELY DETERMINE A CAUSE. THE INSTRUCTIONS FOR USE ADVISE THE USER THAT NEGATIVE PRESSURE IS NEEDED TO MAINTAIN BALLOON DEFLATION. THE APPLICATION OF NEGATIVE PRESSURE WILL ASPIRATE ALL RESIDUAL AIR FROM THE BALLOON AND EASE ENDOSCOPIC ADVANCEMENT. NEGATIVE PRESSURE WILL ALSO AID IN BALLOON PRESERVATION AND OPTIMIZE BALLOON PERFORMANCE. A POSSIBLE CONTRIBUTING FACTOR TO BALLOON MATERIAL FAILURE IS INADEQUATE LUBRICATION OF THE BALLOON WITH A LUBRICATING AGENT. THE INSTRUCTIONS FOR USE DIRECT THE USER TO APPLY A LUBRICATING AGENT TO THE BALLOON TO ALLOW EASIER PASSAGE THROUGH THE ACCESSORY CHANNEL. THIS ACTIVITY WILL AID IN ENDOSCOPIC ADVANCEMENT AND BALLOON PRESERVATION. A BALLOON MATERIAL FAILURE CAN OCCUR IF THE BALLOON IS INFLATED PRIOR TO ADVANCEMENT THROUGH THE ENDOSCOPE OR IF THE BALLOON IS INFLATED WHILE PARTIALLY OR FULLY INSIDE THE ACCESSORY CHANNEL OF THE ENDOSCOPE. THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING PRECAUTION: "DO NOT PRE-INFLATE THE BALLOON". THE INSTRUCTIONS FOR USE STATES, "MAINTAIN BALLOON DEFLATION WITH NEGATIVE PRESSURE AND INTRODUCE INTO THE ACCESSORY CHANNEL OF THE ENDOSCOPE, ADVANCING IN SHORT INCREMENTS UNTIL THE DILATOR IS COMPLETELY VISUALIZED ENDOSCOPICALLY. MONITOR ENDOSCOPICALLY UNTIL THE BALLOON IS IN THE DESIRED POSITION WITHIN THE STRICTURE". THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING WARNING: DURING DELATION DID NOT INFLATE BALLOON BEYOND THE MAXIMUM INDICATED INFLATION PRESSURE AS THIS COULD RESULT IN OVEREXTENSION OR BURSTING OF THE BALLOON. TO ACHIEVE INCREASINGLY LARGE BALLOON DIAMETERS, INCREASE PRESSURE AS INDICATED ON THE CATHETER TAG. ANOTHER POSSIBLE CONTRIBUTING FACTOR IS USING A COMPROMISED INFLATION DEVICE TO INFLATE THE BALLOON. IF THE PRESSURE READING OF THE INFLATION DEVICE IS INACCURATE, THIS COULD CONTRIBUTE TO OVER INFLATION, POSSIBLY RESULTING IN A FAILURE OF THE BALLOON MATERIAL. PRIOR TO DISTRIBUTION, ALL HERCULES 3 STAGE ESOPHAGEAL BALLOON ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TESTING TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MFG REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: CORRECTIVE ACTION IS NOT WARRANTED AT THIS TIME BASED ON THE QUALITY ENGINEERING RISK ASSESSMENT. QA WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.
DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD), A COOK HERCULES 3 STAGE BALLOON ESOPHAGEAL WAS USED. THE BALLOON FAILED TO INFLATE. ANOTHER BALLOON WAS USED TO COMPLETE THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PT'S BODY. THE PT DID NOT REQUIRE ANY ADD'L PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE REPORTER, THE PT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 262063 | HERCULES 3 STAGE BALLOON ESOPHAGEAL | KNQ, DILATOR, ESOPHAGEAL | KNQ | WILSON-COOK MEDICAL INC. | W3320836 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | OLYMPUS ENDOSCOPE (UNK MODEL NUMBER)| COOK SD-60CC-S DILATION SYRINGE |