ANIMAS INSULIN CARTRIDGE
Report
- Report Number
- 2531779-2014-20623
- Event Type
- Malfunction
- Date Received
- July 18, 2014
- Report Date
- July 9, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K032257
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP # 2 DATE OF SUBMISSION 08/07/2014 ¿ CORRECTION: THE ACTUAL DEVICE INVOLVED IN THE INCIDENT HAS NOT BEEN EVALUATED YET. THE PREVIOUSLY REPORTED DEVICE EVALUATION WAS FROM A RETAIN CARTRIDGE SAMPLE WITH THE SAME LOT # AS THE CARTRIDGE INVOLVED IN THE INCIDENT.
FOLLOW-UP # 1 DATE OF SUBMISSION 08/07/2014 ¿ DEVICE EVALUATION: A RETAINED CARTRIDGE SAMPLE WAS EVALUATED BY PRODUCT ANALYSIS (B)(4) 2014 WITH THE FOLLOWING FINDINGS: A RETAIN SAMPLE FROM THE SAME LOT NUMBER WAS TESTED. A VISUAL INSPECTION OF THE CARTRIDGE WAS PERFORMED. NO DAMAGE OR DEFECTS WERE NOTED. A LEAK TEST, FORCE TEST, AND FILL TEST WAS PERFORMED WITH NO FAILURES OBSERVED OR FLUID OBSERVED LEAKING OUT AT THE PLUNGER END OF THE CARTRIDGE. EACH CARTRIDGE LOT IS SUBJECTED TO A STATISTICAL SAMPLING PLAN AND MUST PASS TESTING FOR FORCE (OCCLUSION AND LOSS OF PRIME), CRACKS, AND FOREIGN MATERIAL PRIOR TO RELEASE FOR DISTRIBUTION.
THE DEVICE HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. EACH CARTRIDGE LOT IS SUBJECTED TO A STATISTICAL SAMPLING PLAN AND MUST PASS TESTING FOR FORCE (OCCLUSION AND LOSS OF PRIME), CRACKS, AND FOREIGN MATERIAL PRIOR TO RELEASE FOR DISTRIBUTION.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A SITE/SET/CART (CARTRIDGE LEAK) ISSUE. THE REPORTER ALLEGED THAT THE CARTRIDGE WAS LEAKING AND THAT THE CARTRIDGE WAS WET WHEN REMOVED FROM THE PUMP. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 422685 | ANIMAS INSULIN CARTRIDGE | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | D200037 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 11 YR |