FDA Adverse Event Malfunction Summary report: N

XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM

MDR report key: 3944055 · Received July 18, 2014

Report

Report Number
2024168-2014-04625
Event Type
Malfunction
Date Received
July 18, 2014
Date of Event
June 4, 2014
Report Date
June 23, 2014
Manufacturer
AV-TEMECULA-CT
Product Code
NIQ
PMA / PMN Number
P110019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). CONCOMITANT PRODUCTS: GUIDE WIRE: BALANCE MIDDLEWEIGHT; GUIDE CATH: GUIDING JR 3.5. THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL RELEVANT INFORMATION.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR ANALYSIS. THE STENT DAMAGE AND DIFFICULT TO REMOVE WERE ABLE TO BE CONFIRMED. THE FAILURE TO ADVANCE COULD NOT BE REPLICATED IN A TESTING ENVIRONMENT AS IT WAS BASED ON OPERATIONAL CIRCUMSTANCES. BASED ON A VISUAL, DIMENSIONAL, AND FUNCTIONAL INSPECTION OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. A QUERY/REVIEW OF THE COMPLAINT DATABASE REVEALED NO OTHER SIMILAR INCIDENTS REPORTED FROM THIS LOT. BASED ON THE REVIEWED INFORMATION, NO PRODUCT DEFICIENCY WAS IDENTIFIED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A HEAVILY CALCIFIED LESION IN THE RIGHT CORONARY ARTERY. THE 2.25X15 MM XIENCE XPEDITION STENT DELIVERY SYSTEM (SDS) DID NOT CROSS. DURING RETRACTION OF THE SDS FROM THE ANATOMY, RESISTANCE WAS FELT DURING RETRACTION AT THE MOUTH OF THE GUIDING CATHETER. AFTER RETRACTION IT WAS OBSERVED THAT THERE WAS A FLARED STENT STRUT ON THE PROXIMAL END OF THE STENT. A NON-ABBOTT 3X13 MM STENT WAS ABLE TO CROSS SUCCESSFULLY. NO ADVERSE PATIENT EFFECTS OR CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE WERE REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
422656 XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM DRUG ELUTING CORONARY STENT SYSTEM NIQ AV-TEMECULA-CT 4022441

Patients

Seq Age Sex Outcome Treatment
1