RIATA ST ACTIVE FIXATION
Report
- Report Number
- 2938836-2014-13467
- Event Type
- Injury
- Date Received
- July 18, 2014
- Date of Event
- June 25, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- OTHER
Narratives
NO COMPLAINT WAS RECEIVED WITH THE RETURN OF THE DEVICE. FAILURE EVENT OBSERVED DURING ANALYSIS. FINAL ANALYSIS FOUND A PARTIAL LEAD WITH THE DISTAL TIP RETURNED IN TWO SEGMENTS FOR ANALYSIS. EXTERNAL INSULATION ABRASION WAS NOTED AT 40.0-40.3CM FROM THE DISTAL TIP, CONSISTENT WITH LEAD FRICTION TO ANOTHER DEVICE OR FEATURE OF THE HEART. EXTERNAL INSULATION ABRASIONS WERE NOTED AT 44.3-44.6CM AND 45.0-45.2CM FROM THE DISTAL TIP, CONSISTENT WITH LEAD FRICTION TO THE ICD CAN. INTERNAL INSULATION ABRASIONS WERE NOTED AT 10.9-11.1CM AND 26.9-27.0CM FROM THE DISTAL TIP. THE ETFE COATING WAS INTACT AT THESE LOCATIONS.
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.
ADDITIONAL INFORMATION RECEIVED INDICATED THAT THE LEAD WAS RETURNED.
IT WAS REPORTED THAT THE LEAD WAS EXPLANTED DUE TO AN INSULATION ANOMALY. PATIENT CONDITION IS NORMAL AFTER EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 422968 | RIATA ST ACTIVE FIXATION | DEFIBRILLATION LEAD | NVY | ST. JUDE MEDICAL, INC., CRMD | 7000/65 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR | Required Intervention |