FDA Adverse Event Injury Summary report: N

RIATA ACTIVE FIXATION

MDR report key: 3943998 · Received July 18, 2014

Report

Report Number
2938836-2014-13537
Event Type
Injury
Date Received
July 18, 2014
Date of Event
June 7, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE HIGH VOLTAGE THERAPY DUE TO NOISE ON THE RIGHT VENTRICULAR LEAD. AN INSULATION ANOMALY WAS NOTED ON THE LEAD. THE LEAD WAS CAPPED AND REPLACED. THE PATIENT WAS IN GOOD CONDITION FOLLOWING THE PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
422920 RIATA ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST. JUDE MEDICAL, INC., CRMD 1580/60 NA

Patients

Seq Age Sex Outcome Treatment
1 64 YR Required Intervention