FDA Adverse Event
Malfunction
Summary report: N
RIATA ST ACTIVE FIXATION
MDR report key: 3943987
·
Received July 18, 2014
Report
- Report Number
- 2938836-2014-13466
- Event Type
- Malfunction
- Date Received
- July 18, 2014
- Date of Event
- June 16, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
NO COMPLAINT RECEIVED WITH RETURN OF DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS. EXTERNAL INSULATION ABRASION WAS NOTED AT 8.4-9.0CM FROM THE DISTAL TIP, CONSISTENT WITH LEAD FRICTION TO ANOTHER DEVICE OR FEATURE OF THE HEART. THE ETFE COATING WAS INTACT AT THIS LOCATION. (B)(4). (B)(6).
Description of Event or Problem · 1
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 422481 | RIATA ST ACTIVE FIXATION | DEFIBRILLATION LEAD | NVY | ST. JUDE MEDICAL, INC., CRMD | 7000/65 | NA |