FDA Adverse Event Malfunction Summary report: N

RIATA ST ACTIVE FIXATION

MDR report key: 3943987 · Received July 18, 2014

Report

Report Number
2938836-2014-13466
Event Type
Malfunction
Date Received
July 18, 2014
Date of Event
June 16, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

NO COMPLAINT RECEIVED WITH RETURN OF DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS. EXTERNAL INSULATION ABRASION WAS NOTED AT 8.4-9.0CM FROM THE DISTAL TIP, CONSISTENT WITH LEAD FRICTION TO ANOTHER DEVICE OR FEATURE OF THE HEART. THE ETFE COATING WAS INTACT AT THIS LOCATION. (B)(4). (B)(6).

Description of Event or Problem · 1

THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
422481 RIATA ST ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST. JUDE MEDICAL, INC., CRMD 7000/65 NA

Patients

Seq Age Sex Outcome Treatment
1