FDA Adverse Event
Malfunction
Summary report: N
RIATA ST OPTIM ACTIVE FIXATION
MDR report key: 3943982
·
Received July 18, 2014
Report
- Report Number
- 2938836-2014-13532
- Event Type
- Malfunction
- Date Received
- July 18, 2014
- Date of Event
- June 4, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. NO COMPLAINT RECEIVED WITH RETURN OF DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS. A PARTIAL LEAD WITH THE DISTAL TIP MEASURING 53.0CM WAS RETURNED FOR ANALYSIS. EXTERNAL INSULATION ABRASION WAS NOTED AT 42.8-43.1CM FROM THE DISTAL TIP, CONSISTENT WITH LEAD FRICTION TO ICD CAN. THE ETFE COATING WAS INTACT AT THIS LOCATION. (B)(6).
Description of Event or Problem · 1
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.