RIATA ST ACTIVE FIXATION
MDR report key: 3943975
·
Received July 18, 2014
Report
- Report Number
- 2938836-2014-13435
- Event Type
- Malfunction
- Date Received
- July 18, 2014
- Date of Event
- July 8, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
NO COMPLAINT RECEIVED WITH RETURN OF DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS. EXTERNAL INSULATION ABRASION WAS NOTED AT 12.0-12.2CM FROM THE CONNECTOR PIN, CONSISTENT WITH LEAD FRICTION TO THE ICD CAN. INTERNAL INSULATION ABRASION WAS NOTED AT 13.8-14.1CM FROM THE DISTAL TIP. THE ETFE COATING WAS INTACT AT THESE LOCATIONS. RELIABILITY LABORATORY TECHNICIAN. ST. JUDE MEDICAL (B)(4) RELIABILITY LABORATORY.
Description of Event or Problem · 1
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.