FDA Adverse Event Malfunction Summary report: N

UNIFY QUADRA CRT-D, DF-4 CONNECTOR

MDR report key: 3943897 · Received July 18, 2014

Report

Report Number
2938836-2014-13389
Event Type
Malfunction
Date Received
July 18, 2014
Date of Event
June 17, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NIK
PMA / PMN Number
P030054
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING FOLLOW-UP, POST-PACED T WAVE OVERSENSING ON THE VENTRICULAR CHANNEL WAS OBSERVED. PATIENT WAS ASYMPTOMATIC. PROGRAMMING CHANGES WERE MADE AND RESOLVED THE OVERSENSING. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
421505 UNIFY QUADRA CRT-D, DF-4 CONNECTOR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR NIK ST. JUDE MEDICAL, INC., CRMD CD3249-40Q NA

Patients

Seq Age Sex Outcome Treatment
1 33 YR