FDA Adverse Event Malfunction Summary report: N

VASOVIEW HEMOPRO 2

MDR report key: 3943789 · Received April 30, 2014

Report

Report Number
2242352-2014-00486
Event Type
Malfunction
Date Received
April 30, 2014
Date of Event
April 1, 2014
Report Date
April 3, 2014
Manufacturer
MAQUET CARDIOVASCULAR LLC
Product Code
GEI
PMA / PMN Number
K101274
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SD, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO THE FACTORY FOR EVAL. IT SHOWED SIGNS OF CLINICAL USAGE AND EVIDENCE OF BLOOD. A VISUAL INSPECTION DID NOT IDENTIFY ANY NON-CONFORMITIES. A PRE-CAUTERY TEST WAS PERFORMED ON THE DEVICE WITH A REFERENCE POWER SUPPLY, EXTENSION AND ADAPTER CABLE. THE DEVICE PASSED THE PRE-CAUTERY TEST; IT PRODUCED STEAM AND HEAT DURING SEVERAL ACTIVATIONS AND SHUT OFF WHEN THE TOGGLE WAS RELEASED. A POLYFUSE SUCCESSFULLY SHUT OFF POWER TO THE HEATER AFTER PROLONGED PERIODS OF TIME AND ACTIVATE AFTER 30 SECONDS COOL DOWN. BASED UPON THE EVAL RESULTS, THE REPORTED COMPLAINT COULD NOT BE CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NON CONFORMANCE RECORDED IN THE LOT HISTORY. (B)(4).

Description of Event or Problem · 1

THE HOSP REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, THE VASOVIEW HEMPRO 2 FAILED TO CAUTERIZE WHILE RETRIEVING THE VEIN. THE HOSP DID NOT REPORT ANY PT EFFECTS. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
258277 VASOVIEW HEMOPRO 2 ENDOSCOPIC VESSEL HARVESTING GEI MAQUET CARDIOVASCULAR LLC VH-4000 25089608

Patients

Seq Age Sex Outcome Treatment
1 79 YR