SURESCAN
Report
- Report Number
- 3004209178-2014-13224
- Event Type
- Malfunction
- Date Received
- July 18, 2014
- Date of Event
- June 27, 2014
- Report Date
- June 27, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- LGW
- PMA / PMN Number
- P840001
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
CONCOMITANT MEDICAL PRODUCTS: PRODUCT ID 97740, SERIAL# (B)(4), PRODUCT TYPE: PROGRAMMER, PATIENT; PRODUCT ID 9 77A260, SERIAL# (B)(4), IMPLANTED: (B)(6) 2014, EXPLANTED: (B)(6) 2014, PRODUCT TYPE: LEAD; PRODUCT ID 97754, SERIAL# (B)(4), PRODUCT TYPE: RECHARGER; PRODUCT ID 977A260, SERIAL# (B)(4), IMPLANTED: (B)(6) 2014, PRODUCT TYPE: LEAD; PRODUCT ID NEU_UNKNOWN, LOT# UNKNOWN ANCHOR, PRODUCT TYPE: UNKNOWN; PRODUCT ID NEU_STYLET_ACC, LOT# UNKNOWN, PRODUCT TYPE: ACCESSORY. (B)(4).
IT WAS REPORTED THERE WAS DIFFICULTY ENCOUNTERED WHILE ADVANCING THE LEAD. IT WAS NOTED THAT DURING A REVISION ON (B)(6) 2014 THE HEALTH CARE PROVIDER (HCP) WAS NOT ABLE TO ADVANCE ONE OF THE LEADS. IT WAS REPORTED THEY WERE NOT ABLE TO GET THE STYLET IN THE LEAD. IT WAS NOTED THE STYLET ONLY ¿WENT IN 1-2 INCHES¿. IT WAS REPORTED THAT SINCE THE STYLET WOULD NOT ADVANCE IN THE LEAD THAT IT WAS EXPLANTED AND REPLACED. IT WAS REPORTED AN IMPEDANCE CHECK WAS PERFORMED AND WAS FINE.**INFORMATION OMITTED PERTAINING TO EVENT (B)(4) ¿ ANCHOR AND LEAD MIGRATION. THESE EVENTS ARE NOT REPORTABLE**
ADDITIONAL INFORMATION RECEIVED REPORTED THE MANUFACTURER REPRESENTATIVE SAW THE PATIENT DURING THE WEEK OF THE REPORT AND THEIR THERAPY IS WORKING WELL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 421588 | SURESCAN | STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF | LGW | MEDTRONIC PUERTO RICO OPERATIONS CO. | 97714 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00036 YR |