ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-20601
- Event Type
- Malfunction
- Date Received
- July 18, 2014
- Report Date
- July 10, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP WAS RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/15/2014 WITH THE FOLLOWING FINDINGS: MOISTURE WAS OBSERVED WITHIN THE BATTERY COMPARTMENT, ON THE BATTERY CAP, AND BEHIND THE DISPLAY LENS. THE BATTERY COMPARTMENT AND BATTERY CAP WERE UNDAMAGED AND THE BATTERY CAP WAS ABLE TO SECURE PROPERLY TO THE PUMP. LEAK TESTING FOUND NO PUMP CASE, DISPLAY SCREEN, OR BATTERY COMPARTMENT LEAKS. MOISTURE WAS DISCOVERED ON THE PUMP¿S INTERNAL COMPONENTS.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A CASING/CONDITION (MOISTURE INGRESS) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 421572 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 8 YR |