FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3943696 · Received July 18, 2014

Report

Report Number
2531779-2014-20601
Event Type
Malfunction
Date Received
July 18, 2014
Report Date
July 10, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/15/2014 WITH THE FOLLOWING FINDINGS: MOISTURE WAS OBSERVED WITHIN THE BATTERY COMPARTMENT, ON THE BATTERY CAP, AND BEHIND THE DISPLAY LENS. THE BATTERY COMPARTMENT AND BATTERY CAP WERE UNDAMAGED AND THE BATTERY CAP WAS ABLE TO SECURE PROPERLY TO THE PUMP. LEAK TESTING FOUND NO PUMP CASE, DISPLAY SCREEN, OR BATTERY COMPARTMENT LEAKS. MOISTURE WAS DISCOVERED ON THE PUMP¿S INTERNAL COMPONENTS.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A CASING/CONDITION (MOISTURE INGRESS) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
421572 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 8 YR