FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 3943659
·
Received April 30, 2014
Report
- Report Number
- 1720753-2014-03796
- Event Type
- Malfunction
- Date Received
- April 30, 2014
- Date of Event
- April 10, 2014
- Report Date
- April 30, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE GENERATOR INTERFACE PCB WAS EVALUATED AND REPLACED. A COMPLETE SBC (SINGLE BOARD COMPUTER) REPAIR WAS PERFORMED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM WAS NOT OPERATING CORRECTLY IN SUBTRACTION MODE. A LOSS OF CINE, SUBTRACTION, ROAD-MAPPING, OR OTHER CRITICAL VASCULAR FUNCTIONS COULD RESULT IN DELAY OR TERMINATION/RESCHEDULING OF AN INTERVENTIONAL PROCEDURE. THERE IS NO REPORT OF INJURY OR DEATH.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 258465 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |