FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3943617 · Received July 18, 2014

Report

Report Number
2531779-2014-20590
Event Type
Malfunction
Date Received
July 18, 2014
Report Date
July 9, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 07/09/2014 WITH THE FOLLOWING FINDINGS: DURING TESTING, THE PUMP WAS POWERED ON AND THE DISPLAY SCREEN APPEARED DIM AND DISCOLORED. UNRELATED TO THE DISPLAY ISSUE, TESTING REVEALED THE TIME AND DATE RESET TO FACTORY SETTINGS. THE PUMP CASE WAS REMOVED AND EVALUATION REVEALED THE INTERNAL CLOCK BATTERY ON THE CIRCUIT BOARD HAD FAILED. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED A DISPLAY FAILURE. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 07/09/2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
422876 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 61 YR