BUR 1883672HS 3PK HI SPEED DIAMOND 70DEG
Report
- Report Number
- 3004209178-2014-13218
- Event Type
- Malfunction
- Date Received
- July 18, 2014
- Date of Event
- June 20, 2014
- Report Date
- June 24, 2014
- Manufacturer
- MDT PUERTO RICO OPERATIONS CO
- Product Code
- EQJ
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE PRODUCT ANALYSIS STATED, ¿THE CONDITION OF THE DEVICE SHOWED CUSTOMER USE AS THERE WAS PRESENCE OF CONTAMINANTS ON THE OUTER SHAFT ASSEMBLY¿FROM VISUAL EVALUATION, THE SPIRAL WRAP WAS FOUND HYPEREXTENDED AT THE MIDDLE AND BROKEN NEAR THE TIP. THE DISTAL TIP REMAINED INTACT AND DID NOT DETACH FROM THE BROKEN SPIRAL WRAP. THERE WERE NO SIGNS OF DEVICE MISLOADING BY CUSTOMER AS THERE WERE NO DIMPLES PRESENT ON THE BLUE HUB. THE SPIRAL WRAP HYPEREXTENSION AND BREAKAGE NEAR THE TIP IS INDICATIVE OF POSSIBLE PROCEDURAL / ANATOMICAL / OPERATIONAL FACTORS ENCOUNTERED BY THE CUSTOMER DURING PROCEDURE WHICH MAY HAVE LED TO THE FAILURE. THESE FACTORS INCLUDE AND ARE NOT LIMITED TO DIFFICULT ANATOMICAL LOCATION, HARD BONE/TISSUE STRUCTURE. THESE FACTORS CAN CAUSE THE CUSTOMER TO EXERT PRESSURE, MANIPULATE AND/OR MANEUVER THE DEVICE IN A MANNER THAT MAY LEAD TO ITS BREAKAGE. THE BUR HEAD REMAINED INTACT ON THE BROKEN SPIRAL WRAP PIECE AND DID NOT SHOW ANY ANOMALIES.¿ METHOD: ACTUAL DEVICE EVALUATED; VISUAL INSPECTION; PHOTOGRAPHIC INSPECTION. RESULTS: STRESS PROBLEM. CONCLUSION: OPERATIONAL CONTEXT CAUSED OR CONTRIBUTED TO EVENT. (B)(4).
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
(B)(4). THE PRODUCT ANALYSIS HAS NOT YET BEEN COMPLETED. METHOD: NO TESTING METHODS PERFORMED. RESULTS: RESULTS PENDING COMPLETION OF EVALUATION.
IT WAS REPORTED THAT BLADE BROKE INTER-OPERATIVELY DURING FRONTAL SINUS CASE, NO FRAGMENTS, NO DELAY IN CASE, DOCTOR GRABBED ANOTHER BUR. NO ADDITIONAL INTERVENTION WAS REQUIRED. SHAFT BROKE OFF NEAR TIP WHILE IN SURGICAL FIELD. THIS IS A HIGH-SPEED BUR, AND BREAKS OF HIGH SPEED BURS IN THE PAST HAVE RESULTED IN HEATING UP TO TEMPERATURES HIGH ENOUGH TO CAUSE BURNS. THIS REPORT CONTAINED NO MENTION OF HEATING UP, AND THERE WAS NO INJURY.