FDA Adverse Event Malfunction Summary report: N

TRIAGE BNP TEST FOR BECKMAN COULTER

MDR report key: 3943573 · Received April 30, 2014

Report

Report Number
2027969-2014-00391
Event Type
Malfunction
Date Received
April 30, 2014
Date of Event
April 11, 2014
Report Date
April 11, 2014
Manufacturer
ALERE SAN DIEGO, INC.
Product Code
NBC
PMA / PMN Number
K052789
Removal / Correction Number
2027969-03/20/14-002C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION PENDING.

Description of Event or Problem · 1

CUSTOMER ALLEGING BCI BNP CONTROLS OUT OF RANGE HIGH WITH MICROGENICS CORP CONTROLS L1-3. CUSTOMER USING MICROGENICS CORP CONTROLS - FOR BNP AND CK-MB. ACCOUNT HAS BEEN USING CONTROLS FROM THE SAME COMPANY WITHOUT ANY PREVIOUS ISSUES. CUSTOMER STATES THAT L3 WAS HIGH BUT WITHIN RANGE, HOWEVER, ALSO BECAME OUT OF RANGE AFTER RE-CALIBRATING INSTRUMENT AND RE-RUNNING CONTROLS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
258593 TRIAGE BNP TEST FOR BECKMAN COULTER B-TYPE NATRIURETIC PEPTIDE TEST NBC ALERE SAN DIEGO, INC. 98200 329599

Patients

Seq Age Sex Outcome Treatment
1 MICROGENICS CORP CONTROLS FOR BNP AND CK-MB