FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 3943545 · Received April 30, 2014

Report

Report Number
8020893-2014-01020
Event Type
Malfunction
Date Received
April 30, 2014
Report Date
April 1, 2014
Manufacturer
COVIDIEN, FORMERLY NELLCOR
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NO INFORMATION

Narratives

Additional Manufacturer Narrative · 1

THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) INSPECTED THE DEVICE AND VERIFIED THE MALFUNCTION. THE CSE REPLACED THE GRAPHIC USER INTERFACE (GUI) CENTRAL PROCESSING UNIT (CPU) PRINTED CIRCUIT BOARD (PCB). THE CSE PERFORMED EXTENDED SELF-TESTING ON THE DEVICE AND ALL TESTS PASSED. (B)(4).

Description of Event or Problem · 1

COVIDIEN RECEIVED INFORMATION STATING THAT AN 840 VENTILATOR HAD LINES ACROSS THE UPPER DISPLAY. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME THE MALFUNCTION OCCURRED. THE DATE OF THE EVENT IS UNKNOWN.