FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3943463 · Received July 18, 2014

Report

Report Number
2531779-2014-20580
Event Type
Malfunction
Date Received
July 18, 2014
Report Date
July 9, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/15/2014 WITH THE FOLLOWING FINDINGS: ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. VISUAL INSPECTION OF THE PUMP FOUND SOME COSMETIC DAMAGES. THE BATTERY COMPARTMENT WAS FOUND TO HAVE CRACKED FROM THE THREADS TO THE BASE SEAL AND MOISTURE CORROSION WAS EVIDENT IN THE COMPARTMENT. A LEAK TEST SHOWED A LEAK WHERE THE CRACK WAS LOCATED. THE BATTERY CAP WAS DAMAGED AND A TEST CAP WAS USED IN THE EVALUATION. THE PUMP POWERED UP WITH AUDITORY AND VIBRATORY FEATURES. THE DISPLAY AND ALL THE KEYPAD BUTTONS FUNCTIONED PROPERLY. PUMP CASING WAS OPENED AND NO EVIDENCE OF MOISTURE INTRUSION WAS FOUND INSIDE THE PUMP.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A CASING/CONDITION (MOISTURE INGRESS) ISSUE. REPORTEDLY, MOISTURE WAS OBSERVED IN THE BATTERY COMPARTMENT AND NO DAMAGE TO THE PUMP WAS NOTED. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
422650 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 18 YR