PINNACLE MTL INS NEUT36IDX52OD

MDR report key: 3943343 · Received July 18, 2014

Report

Report Number
1818910-2014-23717
Event Type
Injury
Date Received
July 18, 2014
Date of Event
June 27, 2014
Report Date
August 12, 2014
Manufacturer
DEPUY INTERNATIONAL LTD. 8010379
Product Code
KWA
PMA / PMN Number
PK003523
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PATIENT WAS REVISED TO ADDRESS CHRONIC DISLOCATIONS. METALLOSIS WAS NOTED AND THE CUP REPORTEDLY BOTH OVERLY VERTICAL AND OVERLY ANTEVERTED. DOI: 2010 - DOR: (B)(6) 2014 (RIGHT HIP). THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. A COMPLAINT DATABASE SEARCH FINDS NO OTHER REPORTED INCIDENTS AGAINST THE PROVIDED PRODUCT AND LOT COMBINATIONS. REQUESTS FOR ADDITIONAL INVESTIGATIONAL INPUTS WERE MADE IN ACCORDANCE WITH WI-7915 APPENDIX A. NO ADDITIONAL INFORMATION WAS OBTAINED. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO IDENTIFY ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

PATIENT WAS REVISED TO ADDRESS CHRONIC DISLOCATIONS, METALLOSIS AND AN ANTIVERTED CUP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
423700 PINNACLE MTL INS NEUT36IDX52OD HIP ACETABULAR INSERT/LINER KWA DEPUY INTERNATIONAL LTD. 8010379 2875441

Patients

Seq Age Sex Outcome Treatment
1 40 YR Required Intervention