ANIMAS VIBE
Report
- Report Number
- 2531779-2014-20554
- Event Type
- Malfunction
- Date Received
- July 18, 2014
- Report Date
- July 7, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).
THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/01/2014 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE BLACK BOX AND ALARM HISTORY REVEALED MULTIPLE CALL SERVICE ALARMS. THE PUMP ALARMED WITH A CALL SERVICE ALARM UPON THE FIRST ¿REWIND¿ ATTEMPT. A LANGUAGE CORRUPTION OCCURRED AT A COMPONENT ON THE PRINTED CIRCUIT BOARD RESULTING IN A CALL SERVICE ALARM. THE COMPONENT WAS REPLACED ON THE PCB; THE PUMP WAS POWERED UP AND EXERCISED WITH NO FURTHER ALARMS.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A CALL SERVICE ALARM (CALL SERVICE ALARM ISSUE) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.