ADULT OXYGEN THERAPY BREATHING CIRCUIT
Report
- Report Number
- 9611451-2014-00618
- Event Type
- Death
- Date Received
- July 18, 2014
- Date of Event
- May 7, 2014
- Report Date
- June 24, 2014
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BTT
- PMA / PMN Number
- K983112
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SW
- Reporter Occupation
- OTHER
Narratives
THE RT203 ADULT OXYGEN THERAPY BREATHING CIRCUIT IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT WHICH IS SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K983112. THE COMPLAINT RT203 BREATHING CIRCUIT WAS NOT RETURNED TO FISHER & PAYKEL HEALTHCARE (FPH) (B)(4) FOR EVALUATION AS IT WAS DISCARDED BY THE HOSPITAL. OUR INVESTIGATION IS THEREFORE BASED ON INFORMATION PROVIDED BY THE HOSPITAL. FOLLOWING RECEIPT OF THIS COMPLAINT AN FPH FIELD REPRESENTATIVE VISITED THE HOSPITAL TO GAIN MORE INFORMATION. THE REPRESENTATIVE MET WITH HOSPITAL STAFF AND THE SUPPLIER OF THE MASK. THE HOSPITAL INFORMED THE REPRESENTATIVE THAT THERE WAS NO ISSUE WITH THE RT203. THE RT203 IS MANUFACTURED WITH A 22MM FEMALE CONNECTOR WHICH CONFORMS TO THE STANDARD ISO 5356-1 FOR CONICAL CONNECTORS. FROM THE PHOTOGRAPHS PROVIDED BY THE HOSPITAL, IT APPEARS THE UNOMEDICAL MASK MODEL INVOLVED IN THIS REPORT DOES NOT HAVE THE SAME ISO 5356-1 CONICAL CONNECTOR. THEREFORE THE USE OF THESE TWO PRODUCTS TOGETHER WAS INCOMPATIBLE. THE HOSPITAL IS WORKING WITH THE MASK SUPPLIER TO ADDRESS THIS ISSUE AND CONSIDERING SWITCHING TO A FISHER & PAYKEL HEALTHCARE PRODUCT INSTEAD. AN FPH REPRESENTATIVE IS PROVIDING FURTHER TRAINING AND SUPPORT REGARDING THE CORRECT USE AND COMPATIBILITY OF THE RT203 AND ITS ISO 5356-1 CONICAL CONNECTORS. ALL BREATHING CIRCUITS ARE PRESSURE TESTED FOR LEAKS DURING PRODUCTION AND THOSE THAT FAIL ARE REJECTED. OUR USER INSTRUCTIONS THAT ACCOMPANY THE RT203 ADULT DUAL-HEATED BREATHING CIRCUIT STATE THE FOLLOWING: - "CHECK ALL CONNECTIONS ARE TIGHT BEFORE USE."
A HOSPITAL IN (B)(6) REPORTED VIA THE (B)(6), THAT AN RT203 ADULT OXYGEN THERAPY BREATHING CIRCUIT BECAME DISCONNECTED FROM A UNOMEDICAL AEROSOL ADULT MASK DURING USE. THE PATIENT WAS TERMINALLY ILL DUE TO RESPIRATORY PROBLEMS. IT WAS REPORTED THAT THE PATIENT DIED, POSSIBLY AS A RESULT OF LACK OF OXYGEN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 421523 | ADULT OXYGEN THERAPY BREATHING CIRCUIT | BTT | BTT | FISHER & PAYKEL HEALTHCARE LTD | RT203 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Death | UNOMEDICAL AEROSOL MASK |