ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT
Report
- Report Number
- 9611451-2014-00640
- Event Type
- Malfunction
- Date Received
- July 18, 2014
- Date of Event
- June 10, 2014
- Report Date
- June 20, 2014
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BZE
- PMA / PMN Number
- K983112
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- NURSE
Narratives
(B)(4). METHOD: TWO RT340 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUITS WERE RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) FOR EVALUATION. ONE CIRCUIT WAS FROM LOT 130819 (DATE OF MANUFACTURE: 19 AUGUST 2013) THE OTHER CIRCUIT WAS FROM LOT 130807 (DATE OF MANUFACTURE: 7 AUGUST 2013). THEY WERE VISUALLY INSPECTED AND PRESSURE TESTED FOR LEAK. RESULTS: VISUAL INSPECTION DID NOT IDENTIFY ANY DAMAGE TO THE RETURNED BREATHING CIRCUITS. PRESSURE TEST REVEALED THAT BOTH THE CIRCUITS WERE WITHIN SPECIFICATION. CONCLUSION: WE WERE UNABLE TO DETERMINE WHAT MAY HAVE CAUSED THE LEAK REPORTED BY THE CUSTOMER AS NO FAULT WAS FOUND WITH THE RETURNED BREATHING CIRCUITS. ALL BREATHING CIRCUITS ARE PRESSURE TESTED FOR LEAKS DURING PRODUCTION AND THOSE THAT FAIL ARE REJECTED. THE USER INSTRUCTIONS THAT ACCOMPANY THE RT340 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUIT STATE: "CHECK ALL CONNECTIONS ARE TIGHT BEFORE USE." "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." "SET APPROPRIATE VENTILATOR ALARMS." THE HOSPITAL STAFF CORRECTLY CHECKED THE CIRCUITS PRIOR TO PATIENT USE. THIS IS IN LINE WITH OUR USER INSTRUCTIONS.
(B)(4). THE COMPLAINT DEVICES ARE CURRENTLY EN ROUTE TO FISHER & PAYKEL HEALTHCARE IN (B)(4) FOR EVALUATION. WE WILL PROVIDE A FOLLOW UP REPORT UPON COMPLETION OF OUR INVESTIGATION.
A HOSPITAL IN (B)(6) REPORTED VIA A DISTRIBUTOR THAT TWO RT340 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUITS FAILED THE LEAK TEST PRIOR TO PATIENT USE. THEY FURTHER REPORTED THAT THE EXPIRATORY LIMBS WERE LEAKING.
A HOSPITAL IN (B)(6) REPORTED VIA A DISTRIBUTOR THAT TWO RT340 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUITS FAILED THE LEAK TEST PRIOR TO PATIENT USE. THEY FURTHER REPORTED THAT THE EXPIRATORY LIMBS WERE LEAKING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 421522 | ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT | BZE | BZE | FISHER & PAYKEL HEALTHCARE LTD | RT340 | 130819 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |