530G INSULIN PUMP
Report
- Report Number
- 3004209178-2014-87651
- Event Type
- Malfunction
- Date Received
- July 17, 2014
- Date of Event
- June 16, 2014
- Report Date
- June 17, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZO
- PMA / PMN Number
- P120010
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING, IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.
IT WAS REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR WHILE THE CUSTOMER WAS SITTING ON THE COUCH. THE BLOOD GLUCOSE READING WAS 93 MG/DL. THE CUSTOMER STATED THAT SHE CLEARED THE ALARM BUT THEN IMMEDIATELY THE BATTERY COMPLETELY DEPLETED FROM FULL TO EMPTY ON THE DISPLAY. AN HOUR AFTER THE FIRST MOTOR ERROR ALARM, THE DEVICE ALARMED WITH A SECOND MOTOR ERROR. NO SIGNIFICANT EVENTS LEADING TO THE ALARM WERE OBSERVED. THE CUSTOMER STATED THAT SHE COULD NOT REWIND THE INSULIN PUMP. ADVISED DISCONTINUATION OF THE INSULIN PUMP. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 419511 | 530G INSULIN PUMP | OZO | OZO | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-751NAH |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 35 YR |